HF X-RAY MONOBLOCK FOR ROTATING ANODE TUBE
6kW (XRM6R) / 8kW (XRM8R)
The HF Mobile X-ray Generator consists of an inverter and oil-filled tank which houses the high voltage transformer and tube. It has been designed and developed for OEM application in Radiography for intensive care units. The high frequency output allows for lower kV per exposure, shorter exposure and superior imaging.
TECHNICAL SPECIFICATIONS
| TECHNICAL SPECIFICATIONS | ||
|---|---|---|
| PARAMETER | SPECIFICATION | |
| Input Voltage | 230V ±10% AC, 50Hz, single phase | |
| Model (Maximum Power) |
XRM6R (6kW) 50mA at 120kV 150mA at 40kV |
XRM8R (8kW) 60mA at 120kV 200mA at 40kV |
| HF Generator Frequency | 100kHz | |
| Protections | Over Voltage, Over Current, Over Temperature, Filament Fault, Earth Fault, Rotor Fault, Inverter Fault. | |
| Application | Intensive care units and radiology department | |
| kV Range | 40kV to 120kV (80 steps) | |
| mA Range |
20mA to 150mA (8 steps) |
20mA to 200mA (8 steps) |
| Modes | APR & Manual | |
| mAs | 0.1 to 200mAs in 31 steps | |
| Exposure Time | 10ms to 5sec | |
| Focal Spot | 0.6 / 1.3 | |
| User Interface | Keyboard or Touch screen tablet with microcontroller-based control | |
| Housing |
Monoblock: Contains high voltage transformer,
X-ray tube & filament transformers floating on high voltage. HF Inverter: Contains inverter, control & filament cards, transformers and other interlocks. (Rotor circuitry supplied separately). |
|
| Size (L × W × H) |
Monoblock: 343mm × 185mm × 263mm HF Inverter: 238mm × 236mm × 220mm |
|
| Weight |
Monoblock: 21.5 Kgs HF Inverter: 17 Kgs |
|