3 PHASE HF X-RAY GENERATOR FOR FIXED X-RAY (R/F)

XR15(15kW) / XR20(20kW) / XR32(32kW) / XR40(40kW) / XR50(50kW)/ XR65(65kW)

IONICS offers an extensive range of HF generators that comes in Radiography and Fluroscopy modes. The high frequency output allows for lower kV per exposure, shorter exposure and superior imaging.

TECHNICAL SPECIFICATIONS

TECHNICAL SPECIFICATIONS
PARAMETER SPECIFICATION
Input Voltage Range 380V to 460V AC, 50Hz, 3Phase
Radiographic kV Range 40kV DC to 125kV DC
Maximum Output Power
XR15-II
300mA at 50kV
150mA at 100kV
100mA at 125kV
XR20-II
320mA at 62kV
200mA at 100kV
160mA at 125kV
XR32-II
400mA at 80kV
320mA at 100kV
250mA at 125kV
XR40-II
500mA at 80kV
400mA at 100kV
320mA at 125kV
XR50-II
800mA at 62.5kV
500mA at 100kV
400mA at 125kV
Radiographic mA Range 25mA to 300mA 25mA to 320mA 25mA to 400mA 50mA to 500mA 50mA to 800mA
Radiographic mAs Range 0.25 to 500mAs 0.5 to 500mAs 0.5 to 800mAs
Application Radiography & Fluoroscopy (optional)
Fluoroscopic kV Range 40kV DC to 100kV DC
Fluoroscopic mA Range 0.5 to 5mA
HF Generator Frequency 40kHz
Protections Over Voltage, Over Current, Over Temperature, Earth Fault, Filament Fault, Rotor Fault and Single Phase
AEC Optional
Tube Compatibility Toshiba E7239 / E7242 / (E7252 for 50kW)
Remote Control Touch Screen / Keyboard
Housing (LxWxH) in mm MS cabinet with nylon wheels (432 x 483 x 810; 580 x 420 x 810 for RF unit)
+91 89198 44858 Call Sales
marketing@ionics.co.in Reach out to us @
+91 99853 01016 Call Support