HF X-RAY MONOBLOCK FOR C-ARM MACHINE

3.5kW (XRC3.5) / 5kW (XRC5) / 6kW (XRC6)

The high frequency X-Ray Monoblock consists of an inverter and oil-filled tank which houses the high voltage transformer and tube. It has been designed and developed for OEM application in fluoroscopy and radiography C-arm units. The high radiation output allows for lower kV per exposure, shorter exposure and superior imaging.

TECHNICAL SPECIFICATIONS

TECHNICAL SPECIFICATIONS
PARAMETER XRC3.5 (3.5kW) XRC5 (5kW) XRC6 (6kW)
Input Voltage 230V ±10% AC, 50Hz, single phase
Maximum Output Power 30mA at 110kV
40mA at 80kV
80mA at 40kV
40mA at 120kV
50mA at 100kV
100mA at 50kV
50mA at 120kV
80mA at 75kV
100mA at 60kV
HF Generator Frequency 80kHz
Protections Over Voltage, Over Current, Over Temperature, Filament Fault, Earth Fault, Rotor Fault, Inverter Fault
Application Suitable for Fluoroscopy and Radiography
Fluoroscopic Modes Boosted, ABS, Continuous, Snap shot & Pulsed Boosted, ABS, Continuous, Snap shot & Pulsed Boosted, ABS, Continuous, Snap shot & Pulsed
Fluoroscopic kV Range 40kV to 110kV in 70 steps 40kV to 120kV in 80 steps 40kV to 120kV in 80 steps
Continuous Mode 0.5mA to 3mA 0.5mA to 3mA 0.5mA to 5mA
Single-Pulse & Multi-Pulse, Snap Shot, Boosted Fluoroscopy 6mA and 8mA 10mA and 15mA 10mA and 15mA
Fluoro Timer 0 to 3 min and reset 0 to 3 min and reset 0 to 5 min and reset
Continuous Exposure Time 3 min ON / 8 min OFF
(depends on tube)
3 min ON / 8 min OFF
(depends on tube)
5 min ON / 5 min OFF
(depends on tube)
Radiographic kV Range 40kV to 110kV in 70 steps 40kV to 120kV in 80 steps 40kV to 120kV in 80 steps
Radiographic mA Range 20mA to 80mA 20mA to 100mA 20mA to 100mA
mAs 0.1 to 200mAs in 21 steps
Time 10 ms to 5 s
X-Ray Tube Model Stationary Anode
Dual Focus
Stationary Anode
Dual Focus
Rotating Anode
Dual Focus
Focal Spot 0.6 / 1.5 0.6 / 1.8 0.3 / 0.6
Nominal Voltage 110kV 120kV 120kV
Remote Control Touch screen tablet with Microcontroller based control
Housing – Monoblock Contains high voltage transformer, X-ray tube & filament transformers floating on high voltage
Housing – HF Inverter Contains inverter, control cards, filament card and other interlocks
Size – Monoblock 347mm × 182mm × 211mm 347mm × 182mm × 211mm 347mm × 192mm × 247mm
Size – HF Inverter 233mm × 238mm × 220mm
Weight – Monoblock 17 Kgs 17 Kgs 21.5 Kgs
Weight – HF Inverter 10 Kgs
+91 89198 44858 Call Sales
marketing@ionics.co.in Reach out to us @
+91 99853 01016 Call Support