X-Ray Monoblocks for Orthopaedic

TECHNICAL SPECIFICATIONS

PARAMETER SPECIFICATION
Input Voltage 230V ±10% AC, 50Hz, Single Phase
Output Power
3.5kW 6kW
Maximum Output Voltage
110kV DC 120kV DC
Maximum Output Power
80mA at 44kV 100mA at 60kV
40mA at 80kV 80mA at 75kV
30mA at 110kV 50mA at 120kV
HF Generator Frequency 100kHz
Protections Over Voltage, Over Current, Over Temperature, Filament Fault, Earth Fault, Rotor Fault, Inverter Fault
Application Suitable for Fluoroscopy and Radiography
Fluoroscopy kV Range
40kV to 110kV in 70 steps 40kV to 120kV in 80 steps
Fluoroscopy mA Range
3mA 5mA
Radiography kV Range
40kV to 110kV in 70 steps 40kV to 120kV in 80 steps
Radiography mA Range
20mA to 80mA 20mA to 100mA
mAs 0.1 to 200mAs in 31 steps
Tube Compatibility
X-782 KL-66
Remote Control User Interface Keyboard or Touch screen tablet with microcontroller-based control
Housing Monoblock
HF Inverter
Contains high voltage transformer, X-ray tube & filament transformers floating on high voltage
Contains inverter, control cards, transformers and other interlocks (filament card and transformer supplied separately)
Size (L x W x H) Monoblock
HF Inverter
311mm x 147mm x 211mm 311mm x 156mm x 247mm
238mm x 238mm x 220mm
Weight Monoblock
HF Inverter
17 Kgs 21.5 Kgs
10 Kgs 17 Kgs