HF MONOBLOCK FOR NEONATAL CARE

40W/ 1kW

TECHNICAL SPECIFICATIONS

PARAMETER SPECIFICATION
Rating
40W 1kW
Input Voltage 90 - 264VAC, single phase 47-63Hz (1kW)
Protections Over voltage, over current, over temperature, arc
Interface RS232 & USB
TUBE VOLTAGE PERFORMANCE
Output Voltage
-40kV to -80 kV -40kV to -100 kV
kV Rise
To -80kV in less than 10 msec To -100kV in less than 5 msec
kV Fall time
From -80kV to -25kV in less than 6 msec From -100kV to -25kV in less than 6 msec
kV Repeatability ± 10%
kV Stability ± .5%
kV Ripple ≤ 1% rms
Line Regulation ±0.1% over a range of line voltage from 90 to 264V
Load regulation
±0.1% over a range of emission current from 0.01mA to 0.5mA ±0.1% over a range of emission current from 1.0mA to 10.0mA
kV Accuracy ± 2% of target
BIAS VOLTAGE PERFORMANCE
Bias Voltage Operational Range -300V to 800 V (-300V~0, 0~800V)
V rise 0 to 800V in less than 1.0 msec
V fall time From 800V to 0V in less than 0.02 msec
V repeatability ± 10%
V stability ± 2%
V Ripple ≤ 1% rms
V Accuracy ± 2% of target